Nominal size in mm
110
Accuracy class
±3 %
(±5 % with scale range ≤ 0 ... 125 Pa)
Scale ranges
0 … 50 Pa to 0 … 12,500 Pa
-25 … +25 Pa to -1,500 … +1,500 Pa
further plus/minus ranges on request
Pressure limitation
Steady: Full scale value
Fluctuating: Full scale value
Differential pressure gauge model A2G-15
Overpressure safety
20 kPa
Max. working pressure (static pressure)
20 kPa
Permissible temperature
Ambient: -30 ... +80 °C
Medium: -16 … +50 °C
Temperature effect
When the temperature of the measuring system deviates from the reference temperature (+20 °C): max. ±0.5 %/10 K of full scale value
Ingress protection
IP 54 per EN 60529 / lEC 529
Standard version
Process connection
Plastic, lower mount (LM) or back mount (BM),
G 1/8 (female, in case) for hose connector
Hose connection inner diameter 4 or 6 mm
Standard accessories
3 mounting screws
Straight or angled threaded pressure connection G 1/8 for inner diameter 4 or 6 mm
Compression fitting M12, plastic
Separating diaphragm
Silicone
Movement
Contact-free transmission
Dial
Aluminium, scale angle 90°
Pointer and mark pointer
Plastic
Case
Plastic,
Basic case, case for panel/wall mounting and ring
Window
Makrolon
Electronics
Output signal
0 … 10 V, 3-wire
Power supply UB
DC 15 ... 35 V
Terminal configuration
Permissible max. load
RA > 10 kΩ
Insulation voltage
DC 500 V
Wiring protection
Overvoltage protection: max. DC 35 V
Short-circuit resistance: Sig + to UB -
Reverse voltage protection: UB + to UB -
CE conformity
Pressure equipment directive: 97/23/EC
EMC directive: EN 61000-6-4 und EN 61000-6-2
Relative humidity
45 … 90 % rH, non-condensing
Options
Ingress protection IP 65
Red mark pointer
Internationally recognised traceable calibration certificate
Measuring hose PVC or silicone, inner diameter 4 or 6 mm, length 25 m
Without case
Dimensions in mm
Standard version