Testing standard It's conform IEC61215-2:2016 testing standard requirement
The Purpose It's to determine the ability of the module to withstand thermal mismatch, fatigue and other stresses caused by repeated changes of temperature
The Mainly Specifications
Temperature range
-40oC~85oC
Temperature fluctuation
≤±1oC
Temperature deviation
≤±1oC
Temperature uniformity
≤±2oC
Rise and down speed rate
Satisfy IEC61215 standard requirement
Humidity range
30~98%RH
Relative humidity deviation
±3%FS, deviation 85±2%RH.90±2%RH
Humidity fluctuation
±2%RH
Humidity uniformity
≤2%RH
Thermal testing
module surface's temperature change rate should not more than 100oC/h limit temperature the lowest maintain time 10min, one test is 200 times or 50 cycles
Testing sample quantity
8pcs
Control system
automatic calculus
Display panel
Touch scree
Real-time set display
can show program set time, adjust curve sample rate record cycle, show set vlue curve and actual running valve curve
Real-time curve show
can show program set time, adjust curve sample rate record cycle, show set value curve and actual and humidity value
Communicate interface
RS-485 communicate interface, match on line software can connect to the computer, data save, print and download
Independent control
the program controller will heat up/humidification/dehumidiifcation/cooling etc function complete independent control , save the energy
Language
Chinese and English , can change each other
Power down
reset and power down memory function
Thermal insulation material
High density hard PU foam and glass insulation cotton